CUSTODY AND ROYALTY MEASUREMENT SYSTEM AND EQUIPMENT TOLERANCES

Scope

This Guideline identifies the tolerances and standards that will be considered acceptable performance.

CUSTODY AND ROYALTY MEASUREMENT SYSTEM AND EQUIPMENT TOLERANCES

Frequency
Not Applicable
Procedure

Meter Proving:

Min to Max deviation between trial runs 0.05%
Flow deviation 10% of Flow
Temperature deviation 0.2°F
Trial Runs 10

Meter Factor:

Deviation from current factor to initial factor for propane, butane,  NGL, hexane, ASL, AXL, AL) ± 0.25%
Deviation between current factor and initial factor for AH ± 0.35%

API:

On-line densitometer vs. lab spot sample  : 0.2° API or 0.0010 specific gravity
Densitometer repeatability 0.0005 kg/L
Densitometer calibrated range 0.300 to 1.100 kg/L

Temperature:

Flow Computer vs. Certified
Reference Glass Thermometer
or Electric Temperature Probe 0.2°F (0.1°C)

Pressure:

Flow Computer vs. Reference Gauge ±5 psi

Dead weight tester vs. pressure instrument ±2 psi
Pressure Gauge Accuracy 0.5% of calibrated span
Pressure Transmitter Accuracy ±0.075% of calibrated span.

Pipe Prover:

Sphere Inflation 2% to 4% of pipe prover internal diameter
Base prover volume shift between calibrations. < ±0.02%

Meter Registration Accuracy:

Computer Net vs. Instrument Net 0.005%

Thermometer Accuracy:

Electronic Temperature Probes vs. N.I.S.T. Certified Thermometer  :0.2°F (0.1°C)

Certified Reference Glass Thermometer vs. N.I.S.T. Certified Glass Thermometer  :0.2°F (0.1°C)