SMALL VOLUME PROVER TROUBLESHOOTING FOR OPERATIONS
Problem | Location & Probable Cause | Test/Inspection | Corrective Action |
Poor Repeatability | —– Small Volume Prover —– | ||
Detector(s) | Check detectors for proper operation. |
a) Shield optical switches from ambient light.b) Replace detectors, if faulty.
|
|
Electronics, pulse | Check per instructions in API MPMS Chapter 4, Section 6. |
Adjust, Repair, or replace as required. |
|
Improper plenum pressure |
Verify plenum pressure per manufacturer’s specifications. |
Adjust as required. | |
—– Field Meter —–
|
|||
Meter bearing wear | a) Dismantle and inspect.
b) Analyze pulse train with |
Repair or replace as required. |
|
Damaged PD or turbine meter rotor |
a) Dismantle and inspect.
b) Analyze pulse train with |
Repair or replace as required. |
|
Decrease in meter registration, meter factor increases |
Prover temperature too low. |
Stabilize temperature. | |
Meter | Replace bearings or replace meter. |
||
Increase in meter registration, meter factor decreases |
Prover temperature too high. |
Check thermometer. | |
Fluid viscosity increase |
Stabilize viscosity. | ||
Decrease in meter registration, meter factor decreases |
Build-up on prover pipe wall. |
Clean prover. | |
Increase or decrease in meter registration, meter factor increases or decreases |
Meter calibrator. | Lubricate or replace calibrator. |
|
Severe change in meter factor |
Malfunction in ATC assembly. |
Perform hot and cold test. |
|
Severe decrease in meter factor |
Collapsed bulb and bellows. |
Replace bulb and bellows. |
|
Displacer undersized. | Resize or replace displacer |
||
Sever increase in meter factor |
a) Malfunction of calibrator |
Repair or replace as required. |
|
b) Inner mechanical problem |
b) Inner mechanical problem |